תקציר
We describe two versions of a high temperature flowing afterglow apparatus. With a stainless steel flow tube wrapped with heating tape we have obtained data over the range 300-1300 K. In a version with a ceramic flow tube in a commercial furnace we have obtained data over the range 300-1600 K. The ceramic version is designed to take data up to 1800 K, but we have encountered experimental problems at the upper temperature range. The design modifications to a standard flowing afterglow needed to make measurements at elevated temperatures are described in detail, as are problems associated with operating at elevated temperatures. Samples of data are given.
שפה מקורית | אנגלית |
---|---|
עמודים (מ-עד) | 2142-2148 |
מספר עמודים | 7 |
כתב עת | Review of Scientific Instruments |
כרך | 67 |
מספר גיליון | 6 |
מזהי עצם דיגיטלי (DOIs) | |
סטטוס פרסום | פורסם - יוני 1996 |