In situ measurement of the mass concentration of flame-synthesized nanoparticles using quartz-crystal microbalance

A. Hevroni, H. Golan, A. Fialkov, I. Rahinov, V. Tsionsky, G. Markovich, S. Cheskis

نتاج البحث: نشر في مجلةمقالةمراجعة النظراء

ملخص

A novel in situ method for measurement of mass concentration of nanoparticles (NPs) formed in flames is proposed. In this method, the deposition rate of NPs collected by a molecular beam sampling system is measured by quartz-crystal microbalance (QCM). It is the only existing method which allows direct measurement of NP mass concentration profiles in flames. The feasibility of the method was demonstrated by studying iron oxide NP formation in low-pressure methane/oxygen/nitrogen flames doped with iron pentacarbonyl. The system was tested under fuel-lean and fuel-rich flame conditions. Good agreement between measured QCM deposition rates and their estimations obtained by the transmission electron microscopy analysis of samples collected from the molecular beam has been demonstrated. The sensitivity of the method is comparable to that of particle mass spectrometry (PMS). Combination of the QCM technique with PMS and/or optical measurements can provide new qualitative information which is important for elucidation of the mechanisms governing the NP flame synthesis.

اللغة الأصليةالإنجليزيّة
رقم المقال115102
دوريةMeasurement Science and Technology
مستوى الصوت22
رقم الإصدار11
المعرِّفات الرقمية للأشياء
حالة النشرنُشِر - نوفمبر 2011

بصمة

أدرس بدقة موضوعات البحث “In situ measurement of the mass concentration of flame-synthesized nanoparticles using quartz-crystal microbalance'. فهما يشكلان معًا بصمة فريدة.

قم بذكر هذا